qNMR Experiments
Key Parameters for Quantitation
by Joseph Ray, Naperville
(IL)
- Good signal to noise
- 90° pulse sufficiently
short to excite the entire region uniformly
- Delay between pulses
of at least 5 T1’s
- Flat base line to start
with and baseline correction
- Filter bandwidth set
to at least twice the sw
- Enough data points to
give 15 to 20 points per peak - remember that you lose half the points to
the imaginaries in Fourier transformation
- Acquire
at least 32k and zero fill several times – memory is cheap
- Exception
to 32k for other nuclei requiring 1H decoupling - it is important to limit
the acquisition time to as short as possible to avoid NOE build up during
the acquisition so < 32k may be necessary with zero filling to make up
the difference
- Zero filling should only
be done after dc correction and line broadening otherwise it introduces baseline
artifacts
- Gated decoupling for
non-hydrogen analysis
- Careful phasing on a
spectrum that has been greatly expanded in the vertical direction
- Careful integration
- Place a single integral
over the spectral limits of interest
- Adjust bias and slope
on the full integral
- Cut integral into
smaller regions as close as possible to the peak
- Consistently include
or exclude satellites or account for them when they overlap
See also Joseph
Ray's Anecdotes On Quantitative NMR From The Early 1970ies.
...more under construction...please
bear with us!